MINUTES OF MEETING FOR THE INSPECTION OF PROTOTYPES
OF ALS & BLS AMBULANCES
Reference to Office Order No. SO (P-III) 10-2/2015 dated
23.02.2016, Inspection of prototype of Basic Life Support
(BLS) Ambulance & Advance Life Support (ALS) Ambulance
was held at M/s Toyota Ravi, Multan Road, Lahore dated
03.03.2016. During the inspection of BLS following
distinctions were found;
Reference to Office Order No. SO (P-III) 10-2/2015 dated
28.03.2016, Inspection of prototype of Basic Life Support
(BLS) Ambulance & Advance Life Support (ALS) Ambulance
was held at M/s Ahmed Medix, Lahore dated 30.03.2016.
During the inspection of BLS following distinctions were
found;
PREVIOUS INSPECTION FINDINGS
CURRENT INSPECTION FINDINGS
Out Side Design
i. Name & Logo (design) of institutions to be printed on the
Ambulances should be provided to the M/S Toyota Ravi
motors and they will be pasted on the pre-decided place.
Four Person Seat
i. M/S Toyota Ravi motors was directed to use genuine
seats of Van instead of own made seats (2x2seat) and
make it convenient as per requirement of signed contract.
One Person Crew Seat
i. M/S Toyota Ravi motors was directed to use genuine crew
seats of Van instead of own made seat and make it
convenient as per requirement of signed contract.
Out Side Design
i. Name & Logo (design) of institutions to be printed on the
Ambulances has been requested to the DGHS (Punjab).
(Annexure-A)
Four Person Seat
i. No change made by the firm. Firm is again directed to use
genuine seats of Van instead of its own made seats
(2x2seat) and make it convenient as per requirement of
signed contract. The same was required in the advertised
specifications and further specifically mentioned in the
technical evaluation report as well, which was uploaded on
the health department website. Firm did not objected on
this point at that time. (Annexure-B)
One Person Crew Seat
i. No change made by the firm. Firm is again directed to use
genuine crew seats of Van instead of its own made seat.
The same was required in the advertised specifications
and further specifically mentioned in the technical
evaluation report as well, which was uploaded on the
health department website. Firm did not objected on this
point at that time.
(Annexure-B)

Sign up to view the full document!

lock_open Sign Up
Showing Page:
2/6
2 | P a g e
Main Cabinet
i. Both of the cylinders (48CFT) should be made of USA
Origin and Tested from BOC. Their Certificates are
mandatory. The existing supplied cylinders brand and
origin are not available / mentioned.
ii. M/S Toyota Ravi motors was directed that Cylinder
brackets should be of S.S material instead of any other
material.
iii. Bull nose valve of the cylinder need to be replaced and
should be according to the issued order.
Lights
i. Strobe lights should of red color and at least of power
rating 35watt each, as per requirement of contract. The
existing are in two colors blue and red.
ii. The wiring sizes and standards were not in accordance
with the issued purchase order.
Utility Panel/Console
i. Oxygen flow meter is made of CBMCARE, Italy instead of
Hersill, Spain. M/S Toyota Ravi motors was directed to
make it according to signed contract.
ii. Oxygen Pressure Gauge, Low Oxygen Alarm system and
Main Cabinet
i. No change found. Firm was again unable to provide the
evidence that the supplied Cylinder is of USA Origin. Firm
did not provide any credible evidence which indicates the
brand and origin of cylinder. Firm showed BOC, Pakistan
testing certificate of cylinders which was last year dated
and irrelevant to the cylinders presented before the
committee. (Annexure-C)
ii. Oxygen Cylinder & their piping presented before the
inspection committee were not according to the B.S
Standard color coding. Firm advised to make it as per B.S
Standards for the clear identification of oxygen pipe.
iii. The committee checked and advised the firm to use a non
magnetic stainless steel instead of the existing one, which
will ensure the resistance of rust.
iv. The firm replaced the bull nose valve of cylinder which is
acceptable. The replaced was of O-Two Canada brand.
v. Firm directed to use two switches for two sockets in main
cabinet instead of one switch of two sockets.
Lights
i. Firm did not change the same. Firm was again directed to
make it a single red color strobe light. Moreover, No
documentary evidence provided by the firm and no
physical check /test were performed by the company to
check the power rating of 35 watt of strobe light.
ii. No change made by the company in wiring. The wiring
sizes and standards were not in accordance with the
issued purchase order.
Utility Panel/Console
i. The firm did not replaced the flow meter as per issued
purchase order of Hersil, Spain. Firm was again directed to
make it according to signed contract.
ii. No credible evidence provided by the firm that the

Sign up to view the full document!

lock_open Sign Up
Showing Page:
3/6

Sign up to view the full document!

lock_open Sign Up
End of Preview - Want to read all 6 pages?
Access Now
Unformatted Attachment Preview
MINUTES OF MEETING FOR THE INSPECTION OF PROTOTYPES OF ALS & BLS AMBULANCES Reference to Office Order No. SO (P-III) 10-2/2015 dated 23.02.2016, Inspection of prototype of Basic Life Support (BLS) Ambulance & Advance Life Support (ALS) Ambulance was held at M/s Toyota Ravi, Multan Road, Lahore dated 03.03.2016. During the inspection of BLS following distinctions were found; Reference to Office Order No. SO (P-III) 10-2/2015 dated 28.03.2016, Inspection of prototype of Basic Life Support (BLS) Ambulance & Advance Life Support (ALS) Ambulance was held at M/s Ahmed Medix, Lahore dated 30.03.2016. During the inspection of BLS following distinctions were found; PREVIOUS INSPECTION FINDINGS CURRENT INSPECTION FINDINGS Out Side Design Out Side Design i. Name & Logo (design) of institutions to be printed on the i. Name & Logo (design) of institutions to be printed on the Ambulances should be provided to the M/S Toyota Ravi Ambulances has been requested to the DGHS (Punjab). motors and they will be pasted on the pre-decided place. (Annexure-A) Four Person Seat Four Person Seat i. M/S Toyota Ravi motors was directed to use genuine i. No change made by the firm. Firm is again directed to use seats of Van instead of own made seats (2x2seat) and genuine seats of Van instead of its own made seats make it convenient as per requirement of signed contract. (2x2seat) and make it convenient as per requirement of signed contract. The same was required in the advertised specifications and fur ...